Mass Spectrometer

Location: Federal
Posted: Nov 26, 2025
Due: Dec 11, 2025
Agency: Government of Canada
Type of Government: State & Local
Category:
  • 66 - Instruments and Laboratory Equipment
Solicitation No: 5000085457
Publication URL: To access bid details, please log in.

Description

Trade Agreement:

Canadian Free Trade Agreement (CFTA), the Canada–Chile Free Trade Agreement, the Canada–Colombia Free Trade Agreement, the Canada–Honduras Free Trade Agreement, the Canada–Korea Free Trade Agreement, the Canada–Panama Free Trade Agreement, Canada-Peru Free Trade Agreement, the Comprehensive Economic Free Trade Agreement [CETA], the World Trade Organization – Agreement on Government Procurement [WTO-AGP], the Comprehensive and Progressive Agreement for Trans-Pacific Partnership [CPTPP] and the Canada-Ukraine Free Trade Agreement.

Tendering Procedures:

Request for Proposal to all pre-qualified suppliers on Supply Arrangement E60PV-19EQUI – Laboratory and Scientific Equipment

Competitive Procurement Strategy:

A bid must comply with the requirements of the bid solicitation and meet all mandatory technical evaluation criteria to be declared responsive. The responsive bid with the lowest evaluated price will be recommended for award of a contract.

To be declared responsive, a bid must:

(a)        comply with all the requirements of the bid solicitation; and

(b)        meet all mandatory technical evaluation criteria; and

Bids not meeting (a) or (b) will be declared non-responsive. The responsive bid with the lowest evaluated price will be recommended for award of a contract

Comprehensive Land Claim Agreement: No

Nature of Requirements:

Mitchell Palmer

Team Manager – Procurement – Operations West

Procurement & Contracting Services

Environment and Climate Change Canada

351, boul. Saint-Joseph, Gatineau, QC, K1A 0H3

1-873-499-5126
mitchell.palmer@ec.gc.ca

THIS REQUEST IS RESERVED FOR HOLDERS OF SUPPLY ARRANGEMENT # E60PV-19EQUI ONLY, FOR THE PURCHASE LABORATORY AND SCIENTIFIC EQUIPTMENT.

Only Suppliers currently pre-qualified on Supply Arrangement E60PV-19EQUI have been invited to bid.

As a requirement of the Supply Arrangement, this notice is published on Government Electronic Tendering Service (GETS) for a period of 15 calendar days. The closing date published on this notice identifies how long the notice will be published. For the closing date of any solicitation under the supply arrangement, invited suppliers should refer to the solicitation documents.

Suppliers that do not have a Supply Arrangement for the supply of electrical and electronic products with Public Works and Government Services Canada, cannot submit a bid. Any bids received from suppliers not pre-qualified on the Supply Arrangement will not be evaluated.

Suppliers may qualify under Supply Arrangement E60PV-19EQUI for electrical and electronic products, at any time. Interested suppliers should download solicitation document E60PV-19EQUI from CanadaBuys (SAP Ariba) and submit a response as per the requirements of the Request for Supply Arrangement.

Environment and Climate Change Canada has a requirement for the supply of Gas Chromatograph Mass Spectrometer (GC/MS/MS) Systems for the Pacific & Yukon Laboratory for Environmental Testing (PYLET). The requirement must include all of the following:

Mandatory Technical Criteria:

The following requirements are the mandatory technical evaluation criteria which will be evaluated during the Bid Evaluation. In addition, the Contractor will be required to meet all of the mandatory technical requirements for the duration of the Contract. Bidders are requested to cross reference the mandatory technical criteria in a concise format by using page, paragraph(s) & sub-paragraphs as applicable to their supporting technical documentation.

Item        Criteria        Met/Not Met        Reference to substantiation in the Technical Bid

M1        The GC Subsystem must be an integrated part of the total system and designated for unattended operation. It must be under complete control of the operating/data computer system.

M2        All temperature zones (inlets, oven, detectors, and transfer line to mass spectrometer) must be independently controlled via microprocessors under computer control.

M3        All gas pressures and flows to inlets and detectors electronically controlled via the computer system.

M4        All gas regulators and tubing required for system installation must be included.

M5        The GC method must be part of the full method along with Automatic Sampling parameters, Mass Spectrometer settings, Data Processing options and Reporting Options.

M6        GC mainframe supports up to two additional GC detectors with simultaneous operation with the MS.

M7        The instrument must be configured with a split/splitless inlet.

M8        Oven size capable of accommodating two capillary columns

M9        The GC line must interface to the mass spectrometer (MS/MS) via a direct connection heated transfer line with the effluent of the GC going directly into the MS system. The heating of this connection must be regulated via a microprocessor under computer control with temperature variable from ambient temperature to 350 °C.

M10        The GC must be capable of conducting post-column or mid-point backflushing to be compatible with legacy methods to ensure rapid matrix contaminants purging

M11        The GC must have built in helium saving capabilities and the ability to be upgraded with helium conservation devices that purge the GC and MS with nitrogen when on standby but can switch to helium when needed.

M12        To be compatible with legacy methods, the GC must have capability for doing Dean Switching for ultra-trace analysis of target compounds

M13        To be compatible with legacy methods, the GC must have built-in retention time locking capabilities to utilize the data files acquired in this lab.

4.2.2 GC Autosampler (A/S)

Item        Criteria        Met/Not Met

Reference to substantiation in the Technical Bid

M1        Must be compatible with gas chromatograph and under complete instrument control software.

M2        Autosampler system must provide the capability of simultaneous dual injection of the same sample into the two GC injectors.

4.2.3 Mass Spectrometer (MS/MS)

Item        Criteria        Met/Not Met        Reference to substantiation in the Technical Bid

M1        The MS/MS Subsystem must be an integrated part of the total system and designated for unattended operation and dedicated to performing GC-MS/MS analysis

M2        The mass spectrometer must be of the tandem quadrupole type.

M3        Supplied with appropriate source for Electron Impact Ionization (EI) but capable of being upgraded to Positive and Negative Chemical Ionization (PCI and NCI).

M4        Capable of Scan, Multiple Reaction Monitoring (MRM), and within the same analytical run synchronous Scan and MRM.

M5        Auto-tuning for each of EI, PCI and NCI

M6        Ion source temperature independently heats up to 350 °C

M7        Mass spectra must be spectral library-searchable

M8        MS/MS must be equipped with turbo-molecular pump, allowing quick air-cooling and evacuating the source and the analyzer.

M9        Employs real-time vacuum pressure monitoring of the mass spectrometer.

M10        Fully compatible and can safely operate with hydrogen carrier gas.

M11        To be compatible with legacy methods, the mass spectrometer must have ability to operate in simultaneous MRM/Full Scan mode to collect library searchable spectra while also collecting MRM data for quantitation.

M12        To be compatible with legacy methods, the mass spectrometer must have the ability to switch between helium and nitrogen in the collision cell under instrument control software to be able to perform the pseudoMRM technique.

M13        To be compatible with legacy methods, the instrument must have a technology that automates cleaning of the ion source without the operator present during or after a long sequence.

M14        To be compatible with legacy methods, the mass spectrometer must be capable of heating the quadrupoles to minimum 180°C to ensure solid mass axis stability

4.2.4 Operating/Data Computer System Software

Item        Criteria        Met/Not Met        Reference to substantiation in the Technical Bid

M1        Full control of instrumentation operation including data acquisition and processing.

M2        Multi-tasking and able to acquire and process data in real time.

M3        Full graphical instrument control window with the ability to display system parameters concurrently in real time.

M4        Internal diagnostics including error checking, trouble shooting suggestions, and a complete fault log.

M5        Various calibration modes including external and internal calibration

M6        Multilevel linear and multilevel non-linear calibration capacity.

M7        Ability to insert samples without stopping and restarting a previously started acquisition sequence

M8        Latest NIST MS spectral library included with search capability for qualitative identification

M9        Automatic and manual tuning of MS instruments.

M10        Storage and recall of chromatograms and spectra

M11        Capable of exporting data batched to standard data file types (.txt, .csv, .xls, .mdb) that could be used to upload the data to a LIMS for archiving and final reporting.

M12        Deconvolution software included

M13        Operating software Windows ver. 11.

M14        To be compatible with legacy methods and data files from the last 26 years, the operating system and software much be able to work with existing GCMSMS and GCMS methods and data files on file, a requirement for ISO17025 accreditation.

The current list of standard of operations are: Detection and determination of trace level of volatile polycyclic aromatic hydrocarbons (PAHs) and alkylated polycyclic aromatic hydrocarbons (APAHs); detection and determination of Perchloroethylene; Detection and determination of Trihalomethanes, Detection and determination of personal care and pharmaceutical products (PPCP), Detection, matching, and determination of emergency oil and chemical spills in the environment; Detection and determination of semi-volatile organic pollutants in the surface water; Detection and determination of surfactants and rodenticides in the surface water and other environmental Matrice.

M15        Must be able to generate custom spectral libraries and also import from existing in-house custom spectral libraries.

4.2.5 Operating/Data Computer System Hardware

Item        Criteria        Met/Not Met        Reference to substantiation in the Technical Bid

M1        The PC must feature a processor Intel I7 type processor capable of operating the instrument and processing data smoothly.

M2        Minimum 16 GB RAM

M3        Minimum 500 GB SSD hard drive.

M4        If the instrument requires a dedicated network card for communications, then an additional 10/100/1000 Base-T LAN interface must be provided for access to the laboratory’s network.

M5        The PC must be equipped with a minimum 22” LCD monitor that features minimum resolution of 1280x1024 resolutions

4.2.7 System Implementation and Training

Item        Criteria        Met/Not Met        Reference to substantiation in the Technical Bid

M1        The system must be CSA approved.

M2        The system must be supplied with at least 1 set of consumable parts for initial operation and tool kit necessary for maintaining operation

Show more description Show less description
Contract duration

The estimated contract period will be 3 month(s), with a proposed start date of 2026/01/01.

Trade agreements
  • No trade agreements are applicable to this solicitation process
Reason for limited tendering

A contracting officer can use limited tendering for specific reasons outlined in the applicable trade agreements. The reason for this contract is described below:

  • None
Partner with another business

Are you interested in partnering with other businesses for this opportunity? Add your company to the list of businesses who are interested in partnering .

This list does not replace or affect the tendering procedures for this procurement process. Businesses are still required to respond to bid solicitations, and to compete based on the set criteria. For more information please read the Terms of use .

Daily notification on new contract opportunities

With GovernmentContracts, you can:

  • Find more opportunities and win more business
  • Receive daily alerts for all new bid opportunities
  • Get contract opportunities matched to your business
ONE WEEK FREE TRIAL

See also

...Comprehensive Land Claim Agreement: No Nature of Requirements: ELECTRIC VEHICLE SUPPLY EQUIPMENT E60HN... ...

Government of Canada

Bid Due: 10/15/2026

...and Processing services; and d. STREAM 4: Hydrographic/Bathymetric Consulting services. These... of data ...

Government of Canada

Bid Due: 6/30/2029

...subject to Comprehensive Land Claims Agreements (CLCAs). A Supply Arrangement is established... science ...

Government of Canada

Bid Due: 12/31/2029

...-on Contracts Comprehensive Land Claim Agreement: Yes Nature of Requirements: ONGOING OPPORTUNITY FOR... ...

Government of Canada

Bid Due: 4/16/2046

* Disclaimer: Information regarding bids, requests for proposals (RFPs), or requests for qualifications (RFQs) is provided on this website only for convenience and does not constitute official public notice. Persons wishing to respond to or inquire about bids, RFPs, or RFQs should contact the appropriate government department.