Endoscopy Systems
| Location: |
Maryland |
| Posted: |
Sep 10, 2026 |
| Due: |
Sep 16, 2026 |
| Agency: |
HEALTH AND HUMAN SERVICES, DEPARTMENT OF |
| Type of Government: |
Federal |
| Category: |
- 66 - Instruments and Laboratory Equipment
|
| Solicitation No: |
75F40126Q00381 |
| Publication URL: |
To access bid details, please log in. |
Endoscopy Systems
Active
Contract Opportunity
Department/Ind. Agency
HEALTH AND HUMAN SERVICES, DEPARTMENT OF
Sub-tier
FOOD AND DRUG ADMINISTRATION
Office
FDA OFFICE OF ACQ GRANT SVCS
Looking for contract opportunity help?
General Information
-
Contract Opportunity Type: Combined Synopsis/Solicitation (Original)
-
Original Published Date: Sep 10, 2026 01:33 pm EDT
-
Original Date Offers Due: Sep 16, 2026 03:00 pm EDT
-
Inactive Policy: 15 days after date offers due
-
Original Inactive Date:
Oct 01, 2026
-
Initiative:
Classification
-
Original Set Aside: No Set aside used
-
Product Service Code: 6640 - LABORATORY EQUIPMENT AND SUPPLIES
-
NAICS Code:
-
423490 - Other Professional Equipment and Supplies Merchant Wholesalers
-
Place of Performance:
Silver Spring
,
MD
20993
USA
Description
Under FDA guidance, this OSEL-DIDSR collaboration requires an endoscope to maximize scientific research capacity throughout the center and agency. An endoscope allows OSEL researchers to develop bench testing methods for evaluating the safety and effectiveness of endoscopy devices.
The Government requires an endoscope system (Endoscope) with the following minimum requirements:
-
The Endoscope shall be shipped to the following address:
Attention to: TBD
10903 New Hampshire Ave WO62-3133, US FDA
Silver Spring, MD 20993-
-
Shipping cost is included in the total purchase price.
-
The Endoscope shall include the video processor, colonoscope, medical display, and cart that are interoperable with each other. Once assembled, these components shall operate as a fully functional endoscopy system that can acquire and display live image.
-
The video processor shall satisfy the following requirements:
-
The video processor shall be an FDA-cleared 510(k) medical device. The database is publicly available at 510(k) Premarket Notification.
-
The light source shall be integrated within the video processor.
-
The light source shall use three (3) or more LEDs to synthesize different spectra.
-
The video processor shall have the white light mode.
-
The video processor shall have two (2) or more narrow-band imaging (NBI) modes that utilize different optical spectra generated by the LED-based light source. The NBI modes shall be FDA-cleared for visualizing biological features such as blood vessels or mucosal surfaces.
-
The video processor shall have two (2) or more digital color enhancement modes that utilize different digital image processing techniques. These color enhancement modes shall be FDA-cleared under the FDA product code “PEA”.
-
The video processor shall have the following video output ports and video formats:
-
12G-SDI (3840x2160x60p)
-
3G-SDI (1920x1080x60p)
-
HD-SDI (1900x1080x60i)
-
-
The video processor shall have the capability of saving still images in a standard image file format such as TIFF or JPEG.
-
The video processor shall have the capability of saving video clips in a standard video file format.
-
The colonoscope shall satisfy the following requirements:
-
The colonoscope shall be an FDA-cleared 510(k) colonoscope device.
-
The direction of view shall be forward viewing.
-
The field of view shall be 170 ° or wider.
-
The depth of field shall cover the range of 5-100 mm or wider.
-
The angulation (up/down/right/left) shall be 180°/180°/160°/160° or larger.
-
The colonoscope shall have a water nozzle.
-
The insertion tube diameter shall be 12.8 mm or smaller.
-
The colonoscope shall have an instrument channel with diameter of 3.7 mm or larger.
-
The medical display shall satisfy the following requirements:
-
The signal input ports shall include:
-
12G-SDI
-
3G-SDI
-
Display Port
-
HDMI
-
DVI
-
The pixel count shall be equal to or greater than 3840x2160 pixels (4K).
-
The display size (diagonal) shall be equal to or greater than 32 inches.
-
All components in the system shall be powered by the standard 120 volts, 60 Hertz alternating current power in the U.S.
The Endoscope shall include a total system warranty for at least one (1) year, inclusive of any parts, labor, travel, and shipping cost to return the system to the original equipment manufacturer (OEM) standards.The Contractor shall deliver the endoscope within six (6) weeks of contract award. The Contractor shall provide a one (1)-year warranty period following system acceptance.
Attachments/Links
Contact Information
Contracting Office Address
-
11601 Landsdown St Floor 13
-
Rockville , MD 20852
-
USA
Secondary Point of Contact
History
-
Sep 10, 2026 01:33 pm EDTCombined Synopsis/Solicitation (Original)
Daily notification on new contract opportunities
With GovernmentContracts, you can:
- Find more opportunities and win more business
- Receive daily alerts for all new bid opportunities
- Get contract opportunities matched to your business
ONE WEEK FREE TRIAL
See also
...Follow Global Cardiovascular Research Forum Active Contract Opportunity Notice ID 7436884 Related...
HEALTH AND HUMAN SERVICES, DEPARTMENT OF
Bid Due: 9/22/2026
...Follow Specialized Research Support Services Active Contract Opportunity Notice ID 75N98026Q01186...) intends to ...
HEALTH AND HUMAN SERVICES, DEPARTMENT OF
Bid Due: 9/21/2026
...Follow Sepsis Research Blood Sample Collection Services Active Contract Opportunity Notice ID... SERVICES; ...
DEPT OF DEFENSE
Bid Due: 10/15/2026
...Follow University of Florida Coordinator for Malignant Hyperthermia Research Active Contract...
HEALTH AND HUMAN SERVICES, DEPARTMENT OF
Bid Due: 9/21/2026
* Disclaimer: Information regarding bids, requests for proposals (RFPs), or requests for qualifications (RFQs) is provided on this website only for convenience and does not constitute official public notice. Persons wishing to respond to or inquire about bids, RFPs, or RFQs should contact the appropriate government department.