6515--Non-Invasive Ventilators
| Location: |
Federal |
| Posted: |
Apr 3, 2026 |
| Due: |
Apr 9, 2026 |
| Agency: |
VETERANS AFFAIRS, DEPARTMENT OF |
| Type of Government: |
Federal |
| Category: |
- 65 - Medical, Dental, and Veterinary Equipment and Supplies
|
| Solicitation No: |
36C24726Q0435 |
| Publication URL: |
To access bid details, please log in. |
6515--Non-Invasive Ventilators
Active
Contract Opportunity
Department/Ind. Agency
VETERANS AFFAIRS, DEPARTMENT OF
Sub-tier
VETERANS AFFAIRS, DEPARTMENT OF
Office
247-NETWORK CONTRACT OFFICE 7 (36C247)
Looking for contract opportunity help?
General Information
-
Contract Opportunity Type: Sources Sought (Original)
-
Original Published Date: Apr 03, 2026 04:13 pm EDT
-
Original Response Date: Apr 09, 2026 04:30 pm EDT
-
Inactive Policy: Manual
-
Original Inactive Date:
May 09, 2026
-
Initiative:
Classification
Description
This Request for Information (RFI) is issued solely for information and planning purposes. It does not constitute as a solicitation. All information received in response to this RFI that is marked as proprietary will be handled accordingly. IAW FAR 15.201(e), responses to this notice are not offers and cannot be accepted by the Government to form a binding contract. Responders are solely responsible for all expenses associated with responding to this RFI.
This is NOT a solicitation announcement. This is a sources sought synopsis only. This synopsis aims to gain knowledge of potential qualified sources and their size classification relative to NAICS 339112 (size standard of 1,000 employees). Responses to this synopsis will be used by the Government to make appropriate acquisition decisions. After a review of the responses to this source sought synopsis, a solicitation announcement may be published.
The Department of Veterans Affairs (VA), Regional Procurement Office-RPO East Contracting Office, is seeking sources that can provide Non-Invasive Ventilation System for the Birmingham VA Healthcare System.
The Birmingham VAMC has a requirement for 12 Non-Invasive Ventilators, to include accessories, and staff training. Ventilators should meet the following salient characteristics:
1. Patient Population & Ventilation Range
Must support adult and pediatric patients 12.5 kg.
Must be capable of delivering tidal volumes down to approximately 100 mL.
2. Modes of Ventilation
The ventilator shall provide the following functional therapy modes:
Non Invasive Ventilation (NIV) with pressure support and backup respiratory rate (e.g., Spont PS, S/T or equivalent modes).
Invasive Pressure-Controlled ventilation.
Pressure regulated volume targeted ventilation (PRVC type functionality).
Proportional pressure support or equivalent capabilities to match patient effort.
Integrated high flow oxygen therapy capability.
3. Monitoring Capabilities
Must provide continuous CO and SpO monitoring, including during non invasive ventilation.
Monitoring capabilities must remain available during standby to support patient readiness.
Must store and display at least 72 hours of patient waveforms, parameters, and alarm history.
4. Portability & Power Requirements
Turbine driven ventilator (no external compressor required).
Weight shall not exceed approximately 17 lbs without trolley.
Must include:
Internal backup battery, and
Hot swappable external battery, providing combined runtime of at least 4 hours under typical conditions.
Must be suitable for intrahospital transport while maintaining ventilation.
5. Infection Control Features
Must include dual HEPA filtration or equivalent capability:
One filter protecting external air intake.
One filter on the patient gas pathway/exhalation side.
6. User Interface & Display
Minimum 12 inch color touchscreen or equivalent interface.
Must provide intuitive graphical monitoring and indicators for alarms and patient status.
7. Connectivity
Must support connectivity to common patient monitoring or EMR integration middleware (no brand required). Examples include:
Bedside monitor interface (e.g., via HL7, serial, or proprietary cable systems)
EMR middleware compatibility
8. Regulatory Requirement
Must be a FDA cleared Class II medical device meeting all applicable ventilation safety standards.
9. Accessories (Minimum Required)
Non invasive ventilation masks or interface.
Breathing circuits compatible with both invasive and non invasive modes.
Capnography sensor/interface suitable for NIV use.
Power supply/charger and mounting hardware.
The information identified above is intended to be descriptive, not restrictive, and to indicate the quality of the supplies/services will be satisfactory. It is the responsibility of the interested source to demonstrate to the government that the interested parties can provide the supplies/services that fulfill the required specifications. If you are interested and are capable of providing the sought-out supplies/services, please provide the requested information as well as the information indicated below.
Responses to this notice shall be submitted via email to darius.crane@va.gov no later than 16:30 EST. April 09, 202. Please include in your response company name, address, point of contact, and size of business. Include in the subject line of your email 36C24726Q0435 NON-INVASIVE VENTILATORS . Telephone responses shall not be accepted.
If a solicitation is issued it shall be announced later, and all interested parties must respond to that solicitation announcement separately from the responses to this source sought.
Attachments/Links
Contact Information
Contracting Office Address
-
ONE FREEDOM WAY
-
AUGUSTA , GA 30904
-
USA
Secondary Point of Contact
History
-
Apr 03, 2026 04:13 pm EDTSources Sought (Original)
Daily notification on new contract opportunities
With GovernmentContracts, you can:
- Find more opportunities and win more business
- Receive daily alerts for all new bid opportunities
- Get contract opportunities matched to your business
ONE WEEK FREE TRIAL
See also
...information regarding their current Experience Modification Rate (EMR). This information shall... an offeror ...
VETERANS AFFAIRS, DEPARTMENT OF
Bid Due: 7/03/2026
...for this project o Safety Program (including substance abuse) o Warranty/Guarantee o OSHA ...
Goodwyn Mills Cawood(GMC)
Bid Due: 7/09/2026
...ENVIRONMENTAL VIOLATIONS AND EXPERIENCE MODIFICATION RATE (EMR) See ATTACHMENT 7 All Offerors shall ...
VETERANS AFFAIRS, DEPARTMENT OF
Bid Due: 6/10/2026
...and Documentation of Experience Modification Rate (EMR) (See below instructions). One (1) bid ...
VETERANS AFFAIRS, DEPARTMENT OF
Bid Due: 6/15/2026
* Disclaimer: Information regarding bids, requests for proposals (RFPs), or requests for qualifications (RFQs) is provided on this website only for convenience and does not constitute official public notice. Persons wishing to respond to or inquire about bids, RFPs, or RFQs should contact the appropriate government department.