Sodium Iodide Detector for Thyroid Counting

Location: Massachusetts
Posted: Aug 25, 2026
Due: Aug 31, 2026
Agency: HEALTH AND HUMAN SERVICES, DEPARTMENT OF
Type of Government: Federal
Category:
  • 66 - Instruments and Laboratory Equipment
Solicitation No: 75F40126Q00375
Publication URL: To access bid details, please log in.
Follow
Sodium Iodide Detector for Thyroid Counting
Active
Contract Opportunity
Notice ID
75F40126Q00375
Related Notice
Department/Ind. Agency
HEALTH AND HUMAN SERVICES, DEPARTMENT OF
Sub-tier
FOOD AND DRUG ADMINISTRATION
Office
FDA Office of Acquisitions and Grants Services
Looking for contract opportunity help?

APEX Accelerators are an official government contracting resource for small businesses. Find your local APEX Accelerator (opens in new window) for free government expertise related to contract opportunities.

APEX Accelerators are funded in part through a cooperative agreement with the Department of Defense.

The APEX Accelerators program was formerly known as the Procurement Technical Assistance Program (opens in new window) (PTAP).

General Information
  • Contract Opportunity Type: Combined Synopsis/Solicitation (Original)
  • Original Published Date: Aug 25, 2026 09:58 am EDT
  • Original Date Offers Due: Aug 31, 2026 02:00 pm EDT
  • Inactive Policy: 15 days after date offers due
  • Original Inactive Date: Sep 15, 2026
  • Initiative:
    • None
Classification
  • Original Set Aside: Total Small Business Set-Aside (FAR 19.5)
  • Product Service Code: 6630 - CHEMICAL ANALYSIS INSTRUMENTS
  • NAICS Code:
    • 334516 - Analytical Laboratory Instrument Manufacturing
  • Place of Performance:
    Winchester , MA 01890
    USA
Description

This is a combined synopsis/solicitation for commercial products and commercial services prepared in accordance with the format in FAR 12.6, as supplemented with additional information included in this notice. This announcement constitutes the only solicitation; proposals are being requested, and a written solicitation will not be issued separately.



Solicitation No.: 75F40126Q00375

Contracting Activity: Food and Drug Administration (FDA)

Office/Center: Office of the Commissioner (OC)

Acquisition Type: Commercial Product

Solicitation Type: Request for Quotation (RFQ)

Contract Type: Firm-Fixed-Price Purchase Order

Estimated Value: $20,000

Required Delivery: September 15, 2026



DELIVERABLE:



1. 3X3 LED Temperature stabilized Sodium Iodide Probe compatible with Mirion Osprey Multi Channel Analyzer



Delivery Date: 9/15/2026





2. Digital Tube Based Multichannel Analyzer with USB and Ethernet output



Delivery Date: 09/15/2026



Salient Characteristics



The equipment offered shall meet or exceed all requirements listed below.



Detector Characteristics




  • 3”X 3” NaI(Tl) LED Stabilized Scintillation Detector

  • Energy Range 30 keV to 3 MeV

  • Resolution >7%

  • Available with generic efficiency characterization

  • OPERATING HUMIDITY – 85% non-condensing

  • OPERATING TEMPERATURE – -20 to +50 °C (-4 to 122 °F).

  • Must be compatible and seamlessly compatible with Apex InVivo Software and Mirion Osprey Multichannel Digital Analyzer

  • Diameter less than 3.5”

  • Weight




Multi-Channel Digital Analyzer Characteristics




  • Includes HVPS, Preamplifier and Digital MCA

  • Gain Coarse Gain: x1, x2, x4, x8 software programmable

  • Fine Gain: x1 – x5 software programmable

  • Total Gain: Product of Fine Gain and Coarse Gain, x1 – x40

  • ADC: ADC 14-Bit 20 MHz Sampling

  • ULD/LLD ULD: 0% to 100% of Full Scale software programmable

  • LLD: 0% to 100% of Full Scale software programmable

  • Detector Voltage: HVPS 0 to +1300 V dc: 100 V/s ramping on power up and down

  • Shaping DSP Based Trapezoidal shaping filter operating at 80 MHz

  • Rise Time: 0.2 μs – 5 μs in 200 ns steps. Default: 1 μs for NaI probe

  • Flat Top: 0 to 3 μs in 100 ns steps. Default: 1 μs for NaI probe

  • Digital Spectrum Stabilization

  • Automatic Baseline Restore

  • Automatic Fast Discriminator threshold

  • Seamlessly compatible with Apex InVivo Software

  • USB 2.0 and Ethernet 10/100BaseTX Output

  • 14 pin and 8 pin PMT socket compatible

  • Diameter less than 3.5”

  • Weight
  • ACQUISITION MODES

    • PHA PHA Channels: 256 to 2048 Channels, supports two memory groups of up to 2048 channels each

    • Preset: ROI Counts (Integral), Real or Live Time timers

    • Real and Live Time Resolution: 0.01 s

    • Preset Time: 0 to >4 x 107 s

    • Control: Internal or External Start/Stop control

    • LTC accuracy: 5% (3% typical) up to 50 kcps



  • MCS

    • MCS Dwell range: 1 μs to 999 seconds in 1 μs steps

    • Sweep range: 1 to 232 -1

    • Preset Sweep Counter: 0 to 232 -1 Sweeps (0 implies "Forever")

    • MCS input selection: PHA ROI, external TTL or Fast Discriminator

    • Dwell selection: internal or external

    • MCS Channels: 256 to 2048 Channels, supports two memory groups of up to 2048 channels each



  • 8192 Channels Configurable as two groups of 2048, 1024, 512 or 256 channels for PHA and MCS with 32 bits per channel

  • Integral Non-linearity: ±0.025% of full scale over the top 99% of the selected range

  • Differential Non-linearity: ±1% over the top 99% of the range including the effects from integral non-linearity

  • Gain Drift:
  • Incoming Count Rate (ICR): >250 K cps ICR if not limited by probe/detector

  • Short Circuit Protection

  • High Voltage Power Supply

    • OUTPUT: 0-1300 V at 50 μA max

    • RIPPLE: 1 mV — ACCURACY: 3% Full Scale

    • SETTING RESOLUTION: 14-bit (1/16 384)







WARRANTY AND SUPPORT



The Contractor shall provide a minimum 12-calendar-month warranty covering all parts and labor.



The warranty period shall begin on the date of formal Government acceptance.



Warranty service shall include repair or replacement of defective components necessary to maintain operation of the delivered system.





Review the attached document for complete description of the requirement.




Attachments/Links
Contact Information
Contracting Office Address
  • 5630 Fishers Lane
  • Rockville , MD 20857
  • USA
Primary Point of Contact
Secondary Point of Contact


History
  • Aug 25, 2026 09:58 am EDTCombined Synopsis/Solicitation (Original)
Daily notification on new contract opportunities

With GovernmentContracts, you can:

  • Find more opportunities and win more business
  • Receive daily alerts for all new bid opportunities
  • Get contract opportunities matched to your business
ONE WEEK FREE TRIAL

See also

...Header Information Bid Number: BD-26-2218-BRTAA-BRTAA-132373 Description: ITS Software Procurement...

Commonwealth of Massachusetts

Bid Due: 9/16/2026

...Follow Call for Solutions: Fog Tester Analysis Software Application Active Contract Opportunity... Materials ...

DEPT OF DEFENSE

Bid Due: 9/04/2026

...Follow Call for Solutions: Fog Tester Analysis Software Application Active Contract Opportunity... in ...

DEPT OF DEFENSE

Bid Due: 9/04/2026

...Follow Call for Solutions: Fog Tester Analysis Software Application Active Contract Opportunity... in ...

DEPT OF DEFENSE

Bid Due: 9/04/2026

* Disclaimer: Information regarding bids, requests for proposals (RFPs), or requests for qualifications (RFQs) is provided on this website only for convenience and does not constitute official public notice. Persons wishing to respond to or inquire about bids, RFPs, or RFQs should contact the appropriate government department.