6515--Medical Grade Electrometer for Oncology testing
| Location: |
Oklahoma |
| Posted: |
Aug 20, 2025 |
| Due: |
Sep 3, 2025 |
| Agency: |
VETERANS AFFAIRS, DEPARTMENT OF |
| Type of Government: |
Federal |
| Category: |
- 65 - Medical, Dental, and Veterinary Equipment and Supplies
|
| Solicitation No: |
36C25925Q0746 |
| Publication URL: |
To access bid details, please log in. |
6515--Medical Grade Electrometer for Oncology testing
Active
Contract Opportunity
Department/Ind. Agency
VETERANS AFFAIRS, DEPARTMENT OF
Sub-tier
VETERANS AFFAIRS, DEPARTMENT OF
Office
NETWORK CONTRACT OFFICE 19 (36C259)
Looking for contract opportunity help?
General Information
-
Contract Opportunity Type: Sources Sought (Original)
-
Original Published Date: Aug 20, 2025 03:01 pm MDT
-
Original Response Date: Sep 03, 2025 09:00 am MDT
-
Inactive Policy: Manual
-
Original Inactive Date:
Sep 10, 2025
-
Initiative:
Classification
Description
The Oklahoma City Veterans Affairs Health Care System is seeking sources for a medical electrometer to perform tests performed by the Radiation Oncology department. The device is anticipated to have the following requirements:
Minimum requirements
Remote Access
Full remote access through a web-browser
Flash compatible
Supports your TG-51 QA protocols
Connectivity via WiFi, Ethernet or USB 3.0 Host
Correction Factors
Full remote access through a web-browser
Flash compatible
Supports your TG-51 QA protocols
Connectivity via WiFi, Ethernet or USB 3.0 Host
Library Functions
Extended detector library
Check source library
Machine library
API Functionality
Large touch screen
Desired specifications
Connectivity
Ethernet LAN 1 Gbit/s, Wi-Fi (802.11ac/b/g/n),
USB 3.0 Host
Dimensions (at or near)
249.7mm × 201.9mm × 123.4mm
Current Range
400 fA 24 A
Charge Range
From 24 pC to 65 nC per pulse at a repetition
frequency of 400 Hz
Configurations
TNC Triax, BNC Triax (Adapter)
Zeroing
60 s (default)
Bias Voltage
Â} 50 V Â} 500 V, Â} 2 %
Zero Drift
Â} 0.25 % ( Â} 0.5 %, IEC 60731)
Non-linearity
Â} 0.25 % ( Â} 0.5 %, IEC 60731)
Response Time
Long Term Stability
Â} 0.2 % ( Â} 0.5 % over one year, IEC 60731)
Repeatability
Measurement Modes
Manual, Automatic (beam triggered), Timed
Resolution
0.1 fC
Additional information
This is a request for information and does not represent an order or any obligation on behalf of the VA nor does it commit the VA to a future request for quote (RFQ).
Please provide the country of manufacturing for devices you would consider quoting if an RFQ is issued.
If you consider your company a potential source for this type of device please provide the following information:
Company name,
Name of a point of contact including email address,
Size demographics of company
Questions?
Deadline for questions 8/26
Point of contact: Christopher.delauro@va.gov
Attachments/Links
Contact Information
Contracting Office Address
-
6162 S WILLOW DR SUITE 300
-
Greenwood Village , CO 80111
-
USA
Secondary Point of Contact
History
-
Aug 20, 2025 03:01 pm MDTSources Sought (Original)
Daily notification on new contract opportunities
With GovernmentContracts, you can:
- Find more opportunities and win more business
- Receive daily alerts for all new bid opportunities
- Get contract opportunities matched to your business
ONE WEEK FREE TRIAL
* Disclaimer: Information regarding bids, requests for proposals (RFPs), or requests for qualifications (RFQs) is provided on this website only for convenience and does not constitute official public notice. Persons wishing to respond to or inquire about bids, RFPs, or RFQs should contact the appropriate government department.