Amendment 1to RFQ 1333ND26QNB610465
Clarification questions:
Vendor Question 1:
Could you please confirm the solids content (TSS) of the water? If this is clean water, with less than 2% TSS, please let us know so we can determine the appropriate unit for the application.
NIST Response:
This water is moderately clean, not a slurry (2% TSS is closer to wastewater) but not filtered
Vendor Question 2:
The factory has confirmed that no software is required to configure the meter, as very few ultrasonic meters require software for configuration.
NIST Response:
PC-based data logging is a nice-to-have but not a necessity.
Vendor Question 3: (clarification on the following process conditions for the primary and secondary cooling systems)
Operating temperature of the process fluid (or pipe surface temperature) at the measurement points, for both the primary and secondary cooling loops. This determines whether standard or high-temperature transducers are required.
NIST Response:
No higher than 114 degrees Fahrenheit.
Vendor Question 4: (clarification on the following process conditions for the primary and secondary cooling systems)
Operating pressure of the cooling water at the measurement points (primary and secondary).
NIST Response:
No higher than 10 psia.
Vendor Question 5: (clarification on the following process conditions for the primary and secondary cooling systems)
Confirmation of the process fluid — the SOW references cooling water; please confirm it is treated water with no significant entrained solids or aeration that could affect ultrasonic signal transmission.
NIST Response:
Primary – Heavy water, purified. Secondary - Water with minimal solids. Some algae to be expected but not significant.
As defined in the attached Combined Synopsis Solicitation, Statement of Work and RFO Provisions and Clauses.
| Aug 21, 2026 | [Combined Synopsis/Solicitation (Original)] 7 Ultrasonic Flowmeters |

With GovernmentContracts, you can:
Bid Number and Title Due Date/Time Pre-Bid Conf. Date/Time Purchasing Technician View Documents
Anne Arundel County Public Schools
Bid Due: 9/24/2026
Bid Number: Bid No. RFB-2026-08-14 Bid Title: Kent Ave to N. Oak Utility
La Plata town
Bid Due: 9/24/2026
Follow APC-FY26-000226 FY26 JSD Axon Dedrone Active Contract Opportunity Notice ID 15M10226QA4700237 Related
JUSTICE, DEPARTMENT OF
Bid Due: 9/17/2026
ID: BPM058461 Title: Community Needs Assessment Status: Open Due / Close Date: 9/16/2026
State Government of Maryland
Bid Due: 9/16/2026