Amendment to Solicitation for High Finesse Dual Wavelength Reference Cavity
| Location: |
Maryland |
| Posted: |
May 20, 2025 |
| Due: |
May 27, 2025 |
| Agency: |
COMMERCE, DEPARTMENT OF |
| Type of Government: |
Federal |
| Category: |
- 66 - Instruments and Laboratory Equipment
|
| Solicitation No: |
1333ND25QNB680097 |
| Publication URL: |
To access bid details, please log in. |
Amendment to Solicitation for High Finesse Dual Wavelength Reference Cavity
Active
Contract Opportunity
Notice ID
1333ND25QNB680097
Department/Ind. Agency
COMMERCE, DEPARTMENT OF
General Information View Changes
-
Contract Opportunity Type: Special Notice (Updated)
-
Updated Published Date: May 20, 2025 10:24 am EDT
-
Original Published Date: May 20, 2025 10:15 am EDT
-
Updated Response Date: May 27, 2025 12:00 pm EDT
-
Original Response Date: May 27, 2025 12:00 pm EDT
-
Inactive Policy: 15 days after response date
-
Updated Inactive Date: Jun 11, 2025
-
Original Inactive Date:
Jun 11, 2025
-
Initiative:
Classification
-
Original Set Aside:
-
-
Product Service Code: 6640 - LABORATORY EQUIPMENT AND SUPPLIES
-
NAICS Code:
-
334516 - Analytical Laboratory Instrument Manufacturing
-
Place of Performance:
Gaithersburg
,
MD
20899
USA
Description View Changes
AMENDMENT TO SOLICITATION HIGH FINESSE DUAL WAVELENGTH REFERENCE CAVITY
CLARIFICATION/QUESTIONS AND RESPONSES AND EXTENSION OF RESPONSE TIME
Questions:
Request for more details about the product requirements:
-
The input signal (light): free-space input or fiber input?
-
If free-space input, then the light beam size, collimated or divergent one?
-
For the Photodetectors, any requirements on the speed of the response (bandwidth)? Noise requirement, ….
Response:
A fiber input to the collimator assembly modules is specified in A.12 of the requirements document. Since the modules must be pre-aligned, the fiber input that is used for alignment should remain attached to the system and be included with it. The contractor is responsible for mode matching the light from the fiber into the optical cavity (or cavities) using a set of lenses and configuring additional optics specified in A.12 and A.13 to generate Pound-Drever-Hall locking signals.
The photodetectors must have a bandwidth sufficient for Pound-Drever-Hall locking to the optical reference cavity, at least 10 MHz. The input current noise for the photodetectors should be
Attachments/Links
Contact Information
Contracting Office Address
Secondary Point of Contact
History
Daily notification on new contract opportunities
With GovernmentContracts, you can:
- Find more opportunities and win more business
- Receive daily alerts for all new bid opportunities
- Get contract opportunities matched to your business
ONE WEEK FREE TRIAL
See also
...Beam Supports, or Tubular Steel Supports, Banding of signs to new or existing ...
State Government of Maryland
Bid Due: 6/25/2026
...Scientific (TFS) FEI Helios 660 focused ion beam scanning electron microscope (FIB SEM) ...
COMMERCE, DEPARTMENT OF
Bid Due: 6/18/2026
...accomplished by back-reflection at the far end and subsequent beating with the original ...
COMMERCE, DEPARTMENT OF
Bid Due: 6/22/2026
* Disclaimer: Information regarding bids, requests for proposals (RFPs), or requests for qualifications (RFQs) is provided on this website only for convenience and does not constitute official public notice. Persons wishing to respond to or inquire about bids, RFPs, or RFQs should contact the appropriate government department.