Gas Chromatograph with a Thermal Conductivity Detector and Flame Ioniztion Detector with Methanizer

Location: Maryland
Posted: Sep 17, 2026
Due: Sep 21, 2026
Agency: COMMERCE, DEPARTMENT OF
Type of Government: Federal
Category:
  • 66 - Instruments and Laboratory Equipment
Solicitation No: 1333ND26QNB640601
Publication URL: To access bid details, please log in.
Follow
Gas Chromatograph with a Thermal Conductivity Detector and Flame Ioniztion Detector with Methanizer
Active
Contract Opportunity
Notice ID
1333ND26QNB640601
Related Notice
Department/Ind. Agency
COMMERCE, DEPARTMENT OF
Sub-tier
NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY
Office
DEPT OF COMMERCE NIST
Looking for contract opportunity help?

APEX Accelerators are an official government contracting resource for small businesses. Find your local APEX Accelerator (opens in new window) for free government expertise related to contract opportunities.

APEX Accelerators are funded in part through a cooperative agreement with the Department of Defense.

The APEX Accelerators program was formerly known as the Procurement Technical Assistance Program (opens in new window) (PTAP).

General Information View Changes
  • Contract Opportunity Type: Solicitation (Updated)
  • Updated Published Date: Sep 17, 2026 01:27 pm EDT
  • Original Published Date: Sep 15, 2026 09:03 am EDT
  • Updated Date Offers Due: Sep 21, 2026 12:00 pm EDT
  • Original Date Offers Due: Sep 21, 2026 12:00 pm EDT
  • Inactive Policy: 15 days after date offers due
  • Updated Inactive Date: Oct 06, 2026
  • Original Inactive Date: Oct 06, 2026
  • Initiative:
    • None
Classification
  • Original Set Aside: Total Small Business Set-Aside (FAR 19.5)
  • Product Service Code: 6630 - CHEMICAL ANALYSIS INSTRUMENTS
  • NAICS Code:
    • 334516 - Analytical Laboratory Instrument Manufacturing
  • Place of Performance:
    Gaithersburg , MD 20899
    USA
Description View Changes

***Amendment 0001****



Amendment 0001 provides resposnes to questions received in response to the solicitation.



Q.1. I am reviewing Notice ID 1333ND26QNB640601 and would appreciate some clarification regarding the analytical requirements. Would you be able to share the analytes of interest, the required detection limits, and the expected concentration ranges for the system?



A.1. This analyzer will be used for the analysis of binary gas mixtures in nitrogen or air balance gases. The analyses performed will include a wide range of gaseous standard reference materials including: 1 ppm methane in air and 250 ppm carbon monoxide in nitrogen, both analyzed using the flame ionization detector with methanation, and 0.5 % carbon dioxide in nitrogen analyzed using the thermal conductivity detector. This analyzer will also be used for analyses of ambient level CO in air mixtures, ranging from (40 – 200) ppb.



Q.2. What types of samples will be tested?



A.2. This analyzer will be used for analyses of high pressure gas mixtures in nitrogen or air balance gases.



Q.3. What specific substances or compounds need to be detected or analyzed in these samples?



If available, please also share any specific testing standards or requirements. With this information, we can better understand your application and recommend a suitable instrument configuration and quotation accordingly.



A.3. As stated above, this analyzer will be used for the analysis of high pressure gas mixtures delivered at or below 25 psig. The analyses performed will include a wide range of gaseous standard reference materials including: 1 ppm methane in air and 250 ppm carbon monoxide in nitrogen, both analyzed using the flame ionization detector with methanation, and 0.5 % carbon dioxide in nitrogen analyzed using the thermal conductivity detector.



For improved precision, temperature ramping and stable detector response are required to achieve baseline resolution between components in air balance gas. In all cases the repeatability should be at or below 0.2 % relative area counts.



Q.4. Regarding your software requirements, you cite Visual Basic Version 6.0. Would that software be able to generate an XML file to communicate with our software (CompassCDS)?



A.4. The visual basic software was developed to support communication between the GC software and a 12-port dead-end VICI sampling valve. The software can retrieve data from a txt file generated by the GC software and output the data to excel.



There is not an extension to the solicitation as a result of this amendment.



The question period for this soliciation will close today, September 17,2026, at 4:00 PM ET.



End of Amendment 0001



********



See attached.


Attachments/Links
Contact Information
Contracting Office Address
  • ACQUISITION MANAGEMENT DIVISION 100 BUREAU DR.
  • GAITHERSBURG , MD 20899
  • USA
Primary Point of Contact
Secondary Point of Contact


History

Related Document

Sep 15, 2026[Solicitation (Original)] Gas Chromatograph with a Thermal Conductivity Detector and Flame Ioniztion Detector with Methanizer
Daily notification on new contract opportunities

With GovernmentContracts, you can:

  • Find more opportunities and win more business
  • Receive daily alerts for all new bid opportunities
  • Get contract opportunities matched to your business
ONE WEEK FREE TRIAL

See also

ID: BPM058519 Title: 27-131 AED Maintenance, Inspection and Service Status: Open Due /

State Government of Maryland

Bid Due: 9/30/2026

Follow System Engineering and Technical Assistance (SETA) Support for Public Safety Communications (PSC)

DEPT OF DEFENSE

Bid Due: 9/30/2026

ID: BPM058424 Title: M00P7600920: Cisco ISE Renewal - Addendum II Status: Open Due

State Government of Maryland

Bid Due: 9/30/2026

ID: BPM058543 Title: Home Inspection - King's Reach/Beverly Status: Open Due / Close

State Government of Maryland

Bid Due: 10/05/2026

* Disclaimer: Information regarding bids, requests for proposals (RFPs), or requests for qualifications (RFQs) is provided on this website only for convenience and does not constitute official public notice. Persons wishing to respond to or inquire about bids, RFPs, or RFQs should contact the appropriate government department.