Statement of Requirement
Thyroid Uptake Unit
John J. Pershing VA Medical Center
The Nuclear Medicine service at John J. Pershing VA Medical Center (VAMC) in Poplar Bluff, MO will better serve its Veterans by acquiring this Thyroid Uptake System with Well Counting System with 2 Inch Well Shielding, a multi-purpose spectrum analysis instrument designed for diverse nuclear medicine applications. The system performs Uptake studies, wipe tests, and other user-defined tasks that are accomplished with speed and precision using this fully integrated computer-controlled instrument and its comprehensive selection of application software.
Engineered for mobility, durability and operational efficiency, the unit handles clinical tests, safety compliance tasks, and system administration procedures quickly and accurately. Built-in menus follow conventionally established nuclear medicine procedures, providing step-by-step guidance throughout defined procedures. The system automatically performs calculations, stores patient information and test results, and outputs clear, concise reports. User-defined uptake protocols can be initiated.
SCOPE
REQUIREMENT/SALIENT CHARACTERISTICS/SPECIFICATIONS
Common Nomenclature: MEDICAL SPECTROMETER HARDWARE
Kind of material: ALL IN ONE Touch Screen Computer
Electrical Data:
Power: Low voltage transformer with hospital grade power cord and plug.
Line Voltage:
120 VAC, 60 Hz, 2 AMP circuit breaker
230 VAC, 50 Hz, 1 AMP circuit breaker
Dimensions, size, capacity:
Minimum articulating arm with a minimum of 34 inch vertical range of travel with a minimum of 220 degree collimator rotation.
Principles of operation:
Multi-Channel Analyzer:
Channels: 1024
Spectral Resolution: FWHM 8%
Count Rate: (Maximum) 150,000 cps for Tc-99m
Count Rate Stability: 99%
Gross Count Rate Linearity: Within 5% up to 150,000 cps
Pulse Height Linearity: Within 2% (independent of detector)
Detector High Voltage Adjustment: Automatic adjustment for both probe and well; uses 10 μCi Cs-137 as the calibration source.
Restrictive environmental conditions: User Set: Unlimited user defined Isotopes, setting ROI, half life, name, efficiency and range.
Intended use:
Factory Programmed: Au-198, Ba-133, Co-57, Co-58, Co-60, Cr-51, Cs-137, F-18, Fe-59, Ga-67, Ga-68, Gd-153, Ge-68/Ga-68, Hg-197, I-123, I-125, I-131, In-111, Ir-192, K-42, Lu-177, Na-22, Na-24, Pd-103, Ra-223, Se-75, Sr-85, Tc-99m, Tc-99m/Tl-201, Tl-201, Yb-169, Wide Window.
Equipment with which the item is used, Job related tools:
Probe: 2" x 2" Nal (Tl) integral line scintillation detector
Other pertinent information that describes the item, material or service required:
Uptake Stand:
Dimensions:
Arm down and retracted: 34" l x 26" w x 55" h (86.4 x 66 x 139.7 cm)
Maximum Height: 60" (152.4 cm)
Maximum Length: Arm Extended: 61" (154.9 cm)
Collimated Shield: Flat field collimator meeting IAEA specifications
Positioning: Distance bar and positioning light
Casters: 4" Total locking
Maximum weigh of 300 lbs.
A maximum combined weight with Well Counter: 350 lbs.
Power: Low voltage transformer with hospital grade power cord and plug.
Line Voltage:
120 VAC, 60 Hz, 2 AMP circuit breaker
230 VAC, 50 Hz, 1 AMP circuit breaker
OTHER UNIQUE REQUIREMENTS : Minimum of 2 inch well container
Compatibility with Existing Equipment: TRM Compliance with VA Directives
WINDOWS 10 compatible or Newest Windows Version compliance with VHA standards.
Must be DICOM compliance, Directives or IT Policies, Required.
Coordination with On-going Construction Project (provide the project number): N/A
Warranty: 1 year from date of receipt and verification of installation/operation. Two years part/labor warranty
Training: Initial installation and launch training staff and service support
Service and/or Maintenance: Preventive maintenance and training support, to include two Operating Manuals and two Service Manuals
Safety/ISO/IT/BioMed/SPS & Other Departmental Compliance: Must meet Technical Reference Model (TRM) VA/BioMed Compliance.
DELIVERY
Standard FOB Destination to: John J. Pershing VA Medical Center, 1500 North Westwood Blvd, Poplar Bluff MO 63901.
Government-Furnished Property, Material, Equipment, or Information (GFP, GFM, GFE, or GFI) N/A
SECURITY N/A
SPECIAL MATERIAL REQUIREMENT two Operator’s Manuals and 2 Service Manuals
PLACE OF PERFORMANCE
The system will be installed, serviced, and updated at the facility, John J. Pershing VAMC, Poplar Bluff, MO, Nuclear Medicine Department. Servicing, training, and support of personnel will be per stipulations on the final contract.