Thyroid Uptake System, Poplar Bluff, MO VAMC

Location: Missouri
Posted: Mar 16, 2025
Due: Mar 20, 2025
Agency: VETERANS AFFAIRS, DEPARTMENT OF
Type of Government: Federal
Category:
  • 65 - Medical, Dental, and Veterinary Equipment and Supplies
Solicitation No: 36C25525Q0265
Publication URL: To access bid details, please log in.
Follow
Thyroid Uptake System, Poplar Bluff, MO VAMC
Active
Contract Opportunity
Notice ID
36C25525Q0265
Related Notice
Department/Ind. Agency
VETERANS AFFAIRS, DEPARTMENT OF
Sub-tier
VETERANS AFFAIRS, DEPARTMENT OF
Office
255-NETWORK CONTRACT OFFICE 15 (36C255)
Looking for contract opportunity help?

APEX Accelerators are an official government contracting resource for small businesses. Find your local APEX Accelerator (opens in new window) for free government expertise related to contract opportunities.

APEX Accelerators are funded in part through a cooperative agreement with the Department of Defense.

The APEX Accelerators program was formerly known as the Procurement Technical Assistance Program (opens in new window) (PTAP).

General Information
  • Contract Opportunity Type: Sources Sought (Original)
  • Original Published Date: Mar 14, 2025 11:21 am CDT
  • Original Response Date: Mar 20, 2025 11:00 am CDT
  • Inactive Policy: 15 days after response date
  • Original Inactive Date: Apr 04, 2025
  • Initiative:
    • None
Classification
  • Original Set Aside:
  • Product Service Code: 6515 - MEDICAL AND SURGICAL INSTRUMENTS, EQUIPMENT, AND SUPPLIES
  • NAICS Code:
    • 339113 - Surgical Appliance and Supplies Manufacturing
  • Place of Performance:
    Poplar Bluff , MO 63901
    USA
Description

Statement of Requirement



Thyroid Uptake Unit



John J. Pershing VA Medical Center




  1. GENERAL INFORMATION





The Nuclear Medicine service at John J. Pershing VA Medical Center (VAMC) in Poplar Bluff, MO will better serve its Veterans by acquiring this Thyroid Uptake System with Well Counting System with 2 Inch Well Shielding, a multi-purpose spectrum analysis instrument designed for diverse nuclear medicine applications. The system performs Uptake studies, wipe tests, and other user-defined tasks that are accomplished with speed and precision using this fully integrated computer-controlled instrument and its comprehensive selection of application software.



Engineered for mobility, durability and operational efficiency, the unit handles clinical tests, safety compliance tasks, and system administration procedures quickly and accurately. Built-in menus follow conventionally established nuclear medicine procedures, providing step-by-step guidance throughout defined procedures. The system automatically performs calculations, stores patient information and test results, and outputs clear, concise reports. User-defined uptake protocols can be initiated.



SCOPE




  1. Delivery of the Thyroid Uptake system will allow a bioassay program for quick and efficient measurement of the staff thyroid burden for radiopharmaceuticals, I-123, I-125 and I-131. Measurement results can be reported in cpm, dpm, uCi, or Bq. Action levels may be set by the user for each isotope. System reporting facilitates single bioassay report, summary of multiple bioassays on an individual staff member, and summary report on multiple staff members. Reports are easily printed or saved as PDF files.






  1. The Thyroid Uptake program supports multiple time-stamped uptake measurements, auto decay correction, user-defined uptake protocols, on-screen spectrum acquisition and analysis. The system provides reports that include normal ranges and allow notes/comments entered by the facility physician and technologists.






  1. The wipe test application automatically calculates net contamination levels and reports results in dpm, uCi and Bq. The program allows compliance with regulations for counting and reporting results of wipes taken for contamination surveys. The system allows user to define an unlimited number of “Area” names and wipe “Locations” entered within the areas for testing and documentation. Each area may be designated as “restricted”, “unrestricted”, “package” or “sealed source”. The user may specify count times, activity units, wide window, and individual isotopes and geometric efficiencies or simply utilize the wide range of factory default options.






  1. The program displays and stores the full spectrum with each wipe, automatically calculates net contamination levels and reports results in dpm, uCi or Bq. Wipes above the user defined trigger levels are easily identified. A detailed spectrum analysis may also be performed for any wipe to help identify isotopes causing contamination. Customized reports can be printed or saved in PDF format.






  1. The application specialist will provide training and guidance to ensure safe use of the equipment and the study and administration applications provided. The contractor will provide cybersecurity best practices information and training, when applicable.






  1. The contractor will provide application update to ensure information security in the current equipment and other factors that interconnects with the imaging system including the network, other current peripheral applications and other new tools and software that will be developed or installed to work with the current work station and imaging equipment hardware and software.





REQUIREMENT/SALIENT CHARACTERISTICS/SPECIFICATIONS



Common Nomenclature: MEDICAL SPECTROMETER HARDWARE



Kind of material: ALL IN ONE Touch Screen Computer



Electrical Data:



Power: Low voltage transformer with hospital grade power cord and plug.



Line Voltage:



120 VAC, 60 Hz, 2 AMP circuit breaker



230 VAC, 50 Hz, 1 AMP circuit breaker



Dimensions, size, capacity:



Minimum articulating arm with a minimum of 34 inch vertical range of travel with a minimum of 220 degree collimator rotation.



Principles of operation:





Multi-Channel Analyzer:



Channels: 1024



Spectral Resolution: FWHM 8%



Count Rate: (Maximum) 150,000 cps for Tc-99m



Count Rate Stability: 99%



Gross Count Rate Linearity: Within 5% up to 150,000 cps



Pulse Height Linearity: Within 2% (independent of detector)



Detector High Voltage Adjustment: Automatic adjustment for both probe and well; uses 10 μCi Cs-137 as the calibration source.



Restrictive environmental conditions: User Set: Unlimited user defined Isotopes, setting ROI, half life, name, efficiency and range.



Intended use:



Factory Programmed: Au-198, Ba-133, Co-57, Co-58, Co-60, Cr-51, Cs-137, F-18, Fe-59, Ga-67, Ga-68, Gd-153, Ge-68/Ga-68, Hg-197, I-123, I-125, I-131, In-111, Ir-192, K-42, Lu-177, Na-22, Na-24, Pd-103, Ra-223, Se-75, Sr-85, Tc-99m, Tc-99m/Tl-201, Tl-201, Yb-169, Wide Window.



Equipment with which the item is used, Job related tools:



Probe: 2" x 2" Nal (Tl) integral line scintillation detector



Other pertinent information that describes the item, material or service required:



Uptake Stand:



Dimensions:



Arm down and retracted: 34" l x 26" w x 55" h (86.4 x 66 x 139.7 cm)



Maximum Height: 60" (152.4 cm)



Maximum Length: Arm Extended: 61" (154.9 cm)



Collimated Shield: Flat field collimator meeting IAEA specifications



Positioning: Distance bar and positioning light



Casters: 4" Total locking



Maximum weigh of 300 lbs.



A maximum combined weight with Well Counter: 350 lbs.



Power: Low voltage transformer with hospital grade power cord and plug.



Line Voltage:



120 VAC, 60 Hz, 2 AMP circuit breaker



230 VAC, 50 Hz, 1 AMP circuit breaker



OTHER UNIQUE REQUIREMENTS : Minimum of 2 inch well container



Compatibility with Existing Equipment: TRM Compliance with VA Directives



WINDOWS 10 compatible or Newest Windows Version compliance with VHA standards.





Must be DICOM compliance, Directives or IT Policies, Required.





Coordination with On-going Construction Project (provide the project number): N/A





Warranty: 1 year from date of receipt and verification of installation/operation. Two years part/labor warranty





Training: Initial installation and launch training staff and service support





Service and/or Maintenance: Preventive maintenance and training support, to include two Operating Manuals and two Service Manuals





Safety/ISO/IT/BioMed/SPS & Other Departmental Compliance: Must meet Technical Reference Model (TRM) VA/BioMed Compliance.





DELIVERY




  • Contractor must deliver all components of the system as stated in this solicitation within 30 days of award notice.





Standard FOB Destination to: John J. Pershing VA Medical Center, 1500 North Westwood Blvd, Poplar Bluff MO 63901.





Government-Furnished Property, Material, Equipment, or Information (GFP, GFM, GFE, or GFI) N/A



SECURITY N/A



SPECIAL MATERIAL REQUIREMENT two Operator’s Manuals and 2 Service Manuals



PLACE OF PERFORMANCE



The system will be installed, serviced, and updated at the facility, John J. Pershing VAMC, Poplar Bluff, MO, Nuclear Medicine Department. Servicing, training, and support of personnel will be per stipulations on the final contract.


Attachments/Links
Contact Information
Contracting Office Address
  • VA MEDICAL CENTER 4101 SOUTH 4TH STREET
  • LEAVENWORTH , KS 66048
  • USA
Primary Point of Contact
Secondary Point of Contact


History
  • Mar 14, 2025 11:21 am CDTSources Sought (Original)
Daily notification on new contract opportunities

With GovernmentContracts, you can:

  • Find more opportunities and win more business
  • Receive daily alerts for all new bid opportunities
  • Get contract opportunities matched to your business
ONE WEEK FREE TRIAL
* Disclaimer: Information regarding bids, requests for proposals (RFPs), or requests for qualifications (RFQs) is provided on this website only for convenience and does not constitute official public notice. Persons wishing to respond to or inquire about bids, RFPs, or RFQs should contact the appropriate government department.